Steriliser SF110

 
Contact us for a price
In stock

Standard delivery steriliser

The indispensable safety feature for this hot air steriliser: Setpoint Wait. This means that the programme for sterilisation only starts when the set temperature has been reached. This feature can also be used with freely positionable Pt100 temperature sensors. Here the sterilisation time only begins when the set temperature has been reached at all measurement points, and reliable sterilisation is guaranteed at all times.

Ventilation and Control

  • forced air circulation by quiet air turbine, adjustable in 10 % steps
  • adjustment of pre-heated fresh air admixture by air flap control in 10 % steps
  • vent connection with restrictor flap
  • adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
  • self-diagnostics for fault analysis
  • 1 Pt100 sensor DIN class A in 4-wire-circuit
  • Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
  • parameters adjustable on ControlCOCKPIT: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
  • SetpointWAIT function - the process time does not start until the set temperature is reached
  • AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).
  • Setting of languages on ControlCOCKPIT: German/English/Spanish/French
  • resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED)
  • calibration facility directly at ControlCOCKPIT for three freely selectable temperature values
  • programme stored in case of power failure
  • incl. works calibration certificate for +160°C

Overtemperature protection (double)

  • adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature

Textured Stainless Steel Casing

  • w x h x d: 745 x 864 x 584 mm
  • fully insulated stainless steel door with
    2-point locking (compression door lock)
  • rear zinc-plated steel

Interior - Heating Concept

  • w x h x d: 560 x 480 x 400 mm, 108 l, Depth (less 39 mm for fan)
  • easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
  • 2 stainless steel grids
  • Max. loading of chamber: 175 kg

Temperature Range

  • min.10°C above ambient up to +250°C

Voltage / Power Rating

  • 230 V (+/- 10 %), 50/60 Hz / approx. 2.800 W
  • 115 V (+/- 10 %), 50/60 Hz / approx. 2.200 W

Packing Data

  • net weight approx. 78 kg
  • gross weight carton approx. 103 kg
  • dimensions approx.:
    carton w x h x d: 83 x 105 x 80 cm
  • the appliances must be transported upright

Customs Tariff Number

  • 8419 8998

Country of Origin

  • Federal Republic of Germany

WEEE-Reg.-No.

  • DE 66812464